Model 9110T NOx Analyzer Calibration Procedures
Teledyne Analytical Instruments 216
9.5.1. SETUP ACAL: PROGRAMMING AND AUTO CAL SEQUENCE
To program the example sequence shown in Table 9-4, press:
CONTINUE NEXT PAGE
With STARTING TIME
SAMPLE RANGE = 500.0 PPB NOX=
X
XX.X
< TST TST > CAL CALZ CZLS SETUP
SETUP X.X
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SEQ 1) DISABLED
NEXT MODE EXIT
SETUP X.X SEQ 2) DISABLED
PREV NEXT MODE EXIT
SETUP X.X MODE: DISABLED
NEXT ENTR EXIT
SETUP X.X MODE: ZERO
PREV NEXT ENTR EXIT
SETUP X.X SEQ 2) ZERO
–
SPAN, 1:00:00
PREV NEXT MODE SET EXIT
SETUP X.X TIMER ENABLE: ON
SET> EDIT EXIT
SETUP X.X STARTING DATE: 01–JAN–07
<SET SET> EDIT EXIT
SETUP X.X STARTING DATE: 01–JAN–02
0 4 SEP 0 8 ENTR EXIT
Toggle buttons to set
Day, Month & Year:
Format : DD-MON-YY
SETUP X.X MODE: ZERO–SPAN
PREV NEXT ENTR EXIT