Note If two multi-pod state analyzers are on, the qualifier pop-up menu will
show that only four pattern recognizers are available to each analyzer.
Pattern recognizers a-d and the range recognizer go with the first analyzer
created, and pattern recognizers e-h go with the second analyzer. In the
Full Qualifier Specification pop-up, there will be only one OR gate and
one set of pattern recognizers.
With this Full Qualifier Specification pop-up, you specify a logical
combination of patterns or a range as the qualifier. The pattern recognizers
are always partitioned into the groups of four as shown. Only one
operator is allowed between the patterns in a group. Patterns in
uncomplemented form (a, b, etc.) can only be ORed. The complements of
patterns (a, b, etc.) can only be ANDed. For example, if the first OR
field (gate) is changed to AND, all the patterns for that gate are
complemented, as shown below.
To specify a pattern to be used in the combination, touch the pattern
recognizer field. The field toggles from Off to On and a connection is
drawn from the pattern field to the gate. In figure 5-51, patterns b, c and d
and the range are ORed together, and eandg are ANDed together.
Figure 5-50. Complemented Patterns
HP 16510B Menus
Front-Panel Reference 5-45