HP (Hewlett-Packard) HP 16510B TV Converter Box User Manual


 
These menus show only one pod assigned to each analyzer at power up.
Any number of pods can be assigned to one analyzer, from none to all
five. In the Format menus, only three pods appear at a time in the display.
If there are any pods off screen, an additional field will be present. This
field is labeled Pods .Toviewoff-screenpods,touchthePods field
and rotate the KNOB. The pods are always positioned so that the lowest
numbered pod is on the right and the highest numbered pod is on the left.
Timing and State
Format
Specification
Menu Fields
Seven types of fields are present in the menus. They are:
1) Label
2) Polarity (Pol)
3) Bit assignments
4) Pod threshold
5) Specify Symbols
6) Clock (state analyzer only)
7) Pod Clock (state analyzer only)
8) Clock Period (state analyzer only)
A portion of the menu that is not a field is the Activity Indicators display.
The indicators appear above the bit numbers of each pod. When the logic
analyzer is connected to your target system and the system is running, you
will see in the Activity Indicators display for each channel that has
activity. These tell you that the signals on the channels are transitioning.
The fields in the Format menus are described in the following sections.
The descriptions apply to both the timing and state analyzers unless noted
otherwise.
1Label The label column contains 20 Label fields that you can define. Of the 20
labels, the logic analyzer displays only 8 at one time. To view the labels
that are off screen, rotate the KNOB. The labels roll up and down.
HP 16510B Menus
Front-Panel Reference 5-9