HP (Hewlett-Packard) HP 16510B TV Converter Box User Manual


 
B
Error Messages
This appendix lists the error messages that require corrective action to
restore proper operation of the logic analyzer. There are several messages
that you will see that are merely advisories and are not listed here. For
example, "Load operation complete" is one of these advisories.
The messages are listed in alphabetical order and in bold type.
Autoscale aborted. This message is displayed when the STOP key is
pressed or if a signal is not found 15 seconds after the initiation of
autoscale.
Hardware ERROR: trace point in count block. Indicates the data from
the last acquisition is not reliable and may have been caused by a
hardware problem. Repeat the data acquisition to verify the condition. If
this message re-appears, the logic analyzer requires the attention of service
personnel.
Insufficient memory to load IAL - load aborted. This message indicates
that there is not a block of free memory large enough for the inverse
assembleryouareattemptingtoloadeventhoughtheremaybeenough
memory in several smaller blocks. Try to load the inverse assembler again.
If this load is unsuccessful, load the configuration and the corresponding
inverse assembler separately.
Inverse assembler not loaded--bad object code. Indicates a bad inverse
assembler file on the disc. A new disc or file is required.
Maximum number of symbols already allocated. Indicates an attempt
to create more than 200 symbols.
Maximum of 32 channels per label. Indicates an attempt to assign more
than 32 channels to a label. Reassign channels so that no more than 32 are
assigned to a label.
Must have at least one edge specified. A state clock specification
requires at least one clock edge. This message only occurs if you turn off
all edges in the state clock specification.
HP 16510B Error Messages
Front-Panel Reference B-1