Channel Assignment: Each group of 16 channels (a pod) can be assigned
to Analyzer 1, Analyzer 2, or remain unassigned. The HP 16510B
contains 5 pods.
State Analysis
Memory Data Acquisition: 1024 samples/channel.
Trace
Specification
Clocks: Five clocks are available and can be used by either one or two
state analyzers at any time. Clock edges can be ORed together and operate
in single phase, two phase demultiplexing, or two phase mixed mode.
Clock edge is selectable as positive, negative, or both edges for each clock.
Clock Qualifier: Thehighorlowlevelofuptofourclockscanbe
ANDed with the clock specification. Setup time: 20 ns; hold time: 5 ns.
Pattern Recognizers: Each recognizer is the AND combination of bit (0,
1, or X) patterns in each label. Eight pattern recognizers are available
when one state analyzer is on. Four are available to each analyzer when
two state analyzers are on.
HP 16510B Specifications and Operating Characteristics