Your measurement, then, requires verification of the sequential addresses
the microprocessor looks to and of the data in ROM at these addresses. If
the reset vector fetch is correct (in this example), you will see the
following list of numbers in HEX (default base) when your measurement
results are displayed.
+0000 000000 0000
+0001 000002 04FC
+0002 000004 0000
+0003 000006 8048
+0004 008048 3E7C
This list of numbers will be explained in detail later in this chapter in "The
State Listing."
How Do I
Configure the
Logic Analyzer?
In order to make this state measurement, you must configure the logic
analyzer as a state analyzer. By following these steps you will configure
Analyzer 1 as the state analyzer.
If you are in the State/Timing E Configuration menu you are in the right
place and you can start with step 2; otherwise, start with step 1.
1. Using the field in the upper left corner and the field second from the
left of the display, get the State/Timing E Configuration menu on
screen.
a. Touch the field on the left and when the pop-up appears, touch the
field labeled State/Timing E.
b. Touch the field second from the left. When the pop-up appears,
touch Configuration.
2. In the State/Timing E Configuration menu, change the Analyzer 1
type to State. If Analyzer 1 is already a state analyzer, go on to step
3.
a.TouchthefieldtotherightofType: _________.
b. Touch the field labeled State.
HP 16510B Using the State Analyzer
Front-Panel Reference 8 - 3