State Waveform Menu
Introduction The State Waveform Menu allows you to view state data in the form of
waveformsidentifiedbylabelnameandbitnumber.Upto24waveforms
can be displayed simultaneously. Only state data from the current state
machine can be displayed as waveforms in the State Waveforms menu.
Any intermodule label (i.e., oscilloscope or 1 GHz Timing ) that was
selected when the current machine was a timing analyzer will be deleted
when selecting the State Waveform menu.
The presentation and user interface is generally the same as the Timing
Waveform menu, except the X-axis of the state waveform display
represents only samples, or states instead of seconds. This is true
regardless of whether Count (in the State Trace menu) is set to Time or
Off. As a result, the horizontal axis of the display is scaled by
Samples/Div and Delay in terms of samples from trigger. Marker features
are the same as for State List in that Time or States will only be available
when Count is set to Time or States. The Sample Rate display is not
available in State Waveform even when markers are off.
Accessing the
State Waveform
Menu
The State Waveform menu is accessed by selecting the field directly to the
right of the Module select field in the upper left corner of the screen.
When the pop up appears you will see the options StateWF 1, StateWF 2,
or both depending on which analyzer is a state analyzer. If both analyzers
are state analyzers you will see both StateWF 1 and StateWF 2. You select
your desired option by touching the appropriate field in the pop up.
HP 16510B State Waveform Menu
Front-Panel Reference 10-1