HP (Hewlett-Packard) HP 16510B TV Converter Box User Manual


 
Markers Off When the markers are off they are not displayed, but are still placed at the
specified points in the data. If Stop measurement is on and the Stop
measurement criteria are present in the data, the measurement will stop
even though the markers are off.
Markers Pattern When the markers are set to patterns, you can specify patterns on which
the logic analyzer will place the markers. You can also specify how many
occurrences of each marker pattern the logic analyzer looks for. This use
of the markers allows you to find a specific pattern for each label in the
acquired data.
Patterns for each marker (X and O) can be specified. Patterns can be
specified for both markers in each label. The logic analyzer searches for
thelogical"and"ofpatternsinupto20labels.
In the X (O)-pattern from Trigger field you specify how many occurrences
of the marked pattern from a reference point you want the logic analyzer
to search for. The reference points are:
Trigger
Start (of a trace)
X Marker (only available when searching for the O marker).
Figure 6-12. Markers Pattern
Figure 6-13. Search Reference Pop-up Menu
HP 16510B Interpreting the Display
Front-Panel Reference 6 - 13