The next field to the right of the At____ marker field will pop up when
selected and show you all the labels assigned to the timing analyzer as
shown below.
s/Div (seconds-
per-division)
Field
The seconds-per-division field allows you to change the time window of
the Timing Waveforms menu.
To activate the s/Div field you must touch the field. The field will turn
light blue indicating it can be controlled by the knob. The knob can
increment or decrement the s/Div setting in a 1-2-5 sequence. If you touch
the field again, a pop-up appears. Using the keypad on the pop-up you can
change the seconds-per-division by entering integers and units. The range
for the s/Div field is 10 ns/div to 100 s/div.
When you enter a value from the keypad, the seconds-per-division does
not have to be a 1-2-5 sequence.
Note
Sample period is fixed at 10 ns in the Transitional acquisition mode. In
Glitch mode, changing the s/Div setting changes the sample period for the
next run. To view the sample period after the next run, turn the markers
off if they are on and touch Run.
Figure 6-8. Label Option Pop-up
Interpreting the Display HP 16510B
6 - 8 Front-Panel Reference