HP (Hewlett-Packard) HP 16510B TV Converter Box User Manual


 
Problem Solving
with the
Timing/State
Analyzer
In this example assume you have designed a microprocessor controlled
circuit. You have completed the hardware, and the software designer has
completedthesoftwareandprogrammedtheROM.Whenyouturnyour
circuit on for the first time, your circuit doesn’t work properly. You have
checked the power supply voltages and the system clock and they are
working properly.
Since the circuit has never worked before, you and the software engineer
aren’t sure if it is a hardware or software problem. The problem now
requires some testing to find a solution.
You also notice the circuit fails intermittently. More specifically, it only
fails when the microprocessor attempts to address a routine that starts at
address 8930.
What Am I
Going to
Measure?
To see what might be causing the failure, you decide to start where the
microprocessor goes to the routine that starts at address 8930.
The first thing you check is whether the microprocessor actually addresses
address 8930. The next thing you check is whether the code is correct in
all the steps in this routine.
Your measurement, then, requires verification of:
whether the microprocessor addresses location 8930
whether all the addresses within the routine are correct
whether all the data at the addresses in the routine are correct
If the routine is correct, the state listing will display:
+0000 008930 B03C
+0001 008932 61FA
+0002 008934 67F8
+0003 008936 B03C
+0004 00892E 61FA
Using the Timing/State Analyzer HP 16510B
12-2 Front-Panel Reference