HP (Hewlett-Packard) HP 16510B TV Converter Box User Manual


 
Repetitive
Comparisons
with a Stop
Condition
When you do a comparison in the repetitive trace mode, a stop condition
should be specified. The stop condition is either "Stop Measurement"
when Compare is "Equal," "Not Equal" or "Off." In the case of "Equal",
bits in the compare image must match the corresponding bits in the
acquired data image for the stop condition to be a TRUE. In the case of
"Not Equal", a mismatch on a single bit will cause the stop condition to be
TRUE. When stop conditions are specified in two analyzers, both
analyzers stop when the stop condition of either analyzer is satisfied. It is
an OR function.
The stop measurement function is accessed by selecting the "Specify Stop
Measurement" field found in either the Compare or Difference Listing
menus (see figure 9-4). When this field is selected, the "Stop
Measurement" pop-up appears. The first field in this pop-up, just to the
right of "when," contains either "X-O" or "Compare". When this field is
selected, a pop-up appears in which you select "Compare". When you
select the "Compare" option, you can access and select either the "Equal",
"Not Equal" or "Off" option in the next field to the right.
Figure 9-4. Specify Stop Measurement Field
HP 16510B State Compare Menu
Front-Panel Reference 9-7