In this example, as the state analyzer stores anystate, it will branch to
sequence level 6 if it finds the state given by qualifier e.
The trigger sequence level is used as a boundary for branching between
levels. This level and the levels that occur before it cannot branch to levels
that occur after the trigger level, and vice versa. Therefore, if there are
eight sequence levels and level 5 is the trigger sequence level, then levels
1through5canbranchtolevels1through5only,andlevels6through8
canbranchtolevels6through8only.
You can tell if secondary branch qualifiers have been specified by looking
at the Sequence Levels display. Figure 5-69 shows how the display looks
with the configuration that was given in Figure 5-67. An arrow is drawn
out of level 2 to indicate that branching originates from that level, and an
arrow is drawn into level 4 to indicate that a branch is going into that level.