HP (Hewlett-Packard) HP 8568B TV Converter Box User Manual


 
19. Fast Sweep Time Accuracy Test
(~20
ms)
19. Fast Sweep
Time Accuracy
Test
(430
ms)
Related Adjustment
None
Specification
&lo% for sweep times
5100
seconds
Description
The triangular wave output of a function generator is used to
modulate a 500 MHz signal which is applied to the spectrum analyzer
SIGNAL INPUT. The signal is demodulated in the zero span mode to
display the triangular waveform. Sweep time accuracy for sweep
times
~20
ms is tested by checking the spacing of the signal peaks on
the displayed waveform.
SPECTRUM ANALYZER
SYNTHESIZED SWEEPER
Figure 2-33.
Fast
Sweep Time Accuracy
(~20
ms
Test
Setup)
Equipment
Function Generator
........................................HP
3312A
Universal Counter
..........................................HP
5316A
Signal Generator
...........................................
HP
8340A
Procedure
1. Connect equipment as shown in Figure 2-33.
2. Press
QNSTR
PRESET)
on spectrum analyzer.
3. Key in analyzer settings as follows:
~CENTER
FREQUENCY)
.....................................
500
MHZ
(FREQUENCY SPAN)
........................................100
kHz
4. Set synthesized sweeper for an output frequency of 500 MHz and
an output power level of -10
dBm.
Performance Tests 2-63