6.11.4. RS-485 CONFIGURATION OF COM2
As delivered from the factory, COM2 is configured for RS-232 communications. This port can be re-configured
for operation as a non-isolated, half-duplex RS-485 port capable of supporting up to 32 instruments with a
maximum distance between the host and the furthest instrument being 4000 feet. If you require full-duplex or
isolated operation, please contact Teledyne Instruments Customer Service.
To reconfigure COM2 as an RS-285 port set switch 6 of SW1 to the ON position(see Figure 6-8).
The RS-485 port can be configured with or without a 150 termination resistor. To include the resistor,
install jumper at position JP3 on the CPU board (see Figure 6-8). To configure COM2 as an un-
terminated RS-485 port leave JP3 open.
SW1
Pin 6
CN5
COM2 – RS-485
JP3
CN4
COM2 – RS-232
CN3
COM1 – RS-232
Figure 6-6-8: CPU card Locations of RS-232/486 Switches, Connectors and Jumpers