3.1.2. ELECTRICAL CONNECTIONS
Refer to Figure 3-2 for the location of the rear panel electrical and pneumatic connections.
3.1.2.1. Power Connection
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A current at
your AC voltage and that it is equipped with a functioning earth ground.
CAUTION
CHECK THE VOLTAGE AND FREQUENCY LABEL ON THE REAR PANEL
OF THE INSTRUMENT (SEE FIGURE 3-2 FOR COMPATIBILITY WITH THE
LOCAL POWER BEFORE PLUGGING THE M200EH/EM INTO LINE POWER.
Do not plug in the power cord if the voltage or
frequency is incorrect.
CAUTION
POWER CONNECTION MUST HAVE FUNCTIONING GROUND
CONNECTION.
DO NOT DEFEAT THE GROUND WIRE ON POWER PLUG.
TURN OFF ANALYZER POWER BEFORE DISCONNECTING OR
CONNECTING ELECTRICAL SUBASSEMBLIES.
DO NOT OPERATE WITH COVER OFF.
The M200EH/EM analyzer can be configured for both 100-130 V and 210-240 V at either 50 or 60 Hz., To avoid
damage to your analyzer, make sure that the AC power voltage matches the voltage indicated on the rear panel
serial number label and that the frequency is between 47 and 63 Hz.