Teledyne 200EH/EM TV Converter Box User Manual


 
Operating Instructions Teledyne API - Model 200EH/EM Operation Manual
82
6.9. SETUP PASS: PASSWORD FEATURE
The M200EH/EM provides password protection of the calibration and setup functions to prevent unauthorized
adjustments. When the passwords have been enabled in the
PASS menu item, the system will prompt the user
for a password anytime a password-protected function is requested.
There are three levels of password protection, which correspond to operator, maintenance, and configuration
functions. Each level allows access to all of the functions in the previous level.
Table 6-10: Password Levels
PASSWORD LEVEL MENU ACCESS ALLOWED
No password Operator TEST, MSG, CLR
101 Maintenance CAL, CALZ, CALS
818 Configuration SETUP, VARS, DIAG
To enable or disable passwords, press the following keystroke sequence:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
Toggle this
button to
enable, disable
password
feasture
SETUP X.X PASSWORD ENABLE: OFF
OFF ENTR EXIT
SETUP X.X PASSWORD ENABLE: ON
ON ENTR EXIT
Example: If all passwords are enabled, the following keypad sequence would be required to enter the SETUP
menu:
04521C (DCN5731)