Teledyne 200EH/EM TV Converter Box User Manual


 
Teledyne API - Model 200EH/EM Operation Manual Specifications, Approvals and Warranty
3
2. SPECIFICATIONS, APPROVALS AND WARRANTY
2.1. M200EH/EM OPERATING SPECIFICATIONS
Table 2-1: Model 200EH/EM Basic Unit Specifications
Min/Max Range
(Physical Analog Output)
200EH: Min: 0-5 ppm; Max: 0-5000 ppm
200EM: Min: 0-1 ppm; Max: 0-200 ppm
Measurement Units ppm, mg/m
3
(user selectable)
Zero Noise <20 ppb (RMS)
Span Noise <0.2% of reading above 20 ppm
Lower Detectable Limit 40 ppb (2x noise as per USEPA)
Zero Drift (24 hours) <20 ppb (at constant temperature and voltage.)
Zero Drift (7 days) <20 ppb (at constant temperature and voltage.)
Span Drift (7 Days) <1% of reading (at constant temperature and voltage.)
Linearity 1% of full scale
Precision 0.5% of reading
Lag Time
20 s
Rise/Fall Time 95% in <60 s (~10 s in NO only or NO
X
only modes)
Gas Flow Rates
200EH: -40 cm³/min sample gas through NO
2
converter & sensor module
-250 cm
3
/min ± 10% though bypass manifold;
-290 cm³/min total flow
200EM: -250 cm³/min sample gas through NO
2
converter & sensor module
O
2
Sensor option adds 80 cm³/min to total flow though M200EH/EM when installed;
Temperature Range
5 - 40 C operating range
Humidity Range 0-95% RH non-condensing
Dimensions H x W x D 18 cm x 43 cm x 61 cm (7" x 17" x 23.6")
Weight, Analyzer 18 kg (40 lbs) without options
Weight, Ext Pump Pack 7 kg (16 lbs)
AC Power Rating
100 V, 50/60 Hz (3.25A);
115 V, 60 Hz (3.0 A);
220 - 240 V, 50/60 Hz (2.5 A)
Power, Ext Pump
100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A);
220 - 240 V, 50/60 Hz (2.5 A)
Environmental Installation category (over-voltage category) II; Pollution degree 2
Analog Outputs 4 user configurable outputs
Analog Output Ranges
All Outputs: 0.1 V, 1 V, 5 V or 10 V
Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over-range
Analog Output Resolution 1 part in 4096 of selected full-scale voltage (12 bit)
Status Outputs 8 Status outputs from opto-isolators, 7 defined, 1 spare
Control Inputs 6 Control inputs, 4 defined, 2 spare
Alarm outputs
2 relay alarms outputs (Optional equipment) with user settable alarm limits
- 1 Form C: SPDT; 3 Amp @ 125 VAC
Serial I/O
1x RS-232; 1x RS-485 or RS-232 (configurable)
Communication speed: 300 - 115200 baud (user selectable)
Certifications
CE: EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A,
ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.
04521C (DCN5731)