6.13.5.8. Turning an Analog Output On or Off
Each output can be temporarily turned off. When off no data is sent to the output. Electronically, it is still active,
there is simply no data so the signal level at the rear of the instrument will fall to zero.
To turn an individual analog output channel ON/OFF, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_2 OUTPUT: ON
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_2 OUTPUT: 5V
SET> EDIT EXIT
DIAG AIO DATA_OUT_2 OUTPUT: ON
ON ENTR EXIT
DIAG AIO DATA_OUT_2 OUTPUT: OFF
OFF ENTR EXIT
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
Toggle this key
to turn the
channel
ON/OFF
Continue pressing SET> until ...
Continue pressing SET> until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT
04521C (DCN5731)