Operating Instructions Teledyne API - Model 200EH/EM Operation Manual
136
6.13.7.3. Electrical Test
The electrical test function creates a current, which substitutes the PMT signal, and feeds it into the preamplifier
board. This signal is generated by circuitry on the pre-amplifier board itself and tests the filtering and
amplification functions of that assembly along with the A/D converter on the motherboard. It does not test the
PMT itself. The electrical test should produce a PMT signal of about 2000 ±1000 mV.
To activate the electrical test press the following keys.
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
DIAG SIGNAL I / O
PREV
NEXT
JUMP
ENTR
EXIT
DIAG ELEC A1:NXCNC1=100PPM NOX=XXX.X
<TST TST> EXIT
Press NEXT until…
DIAG ELECTRICAL TEST
PREV NEXT
ENTR
EXIT
Press
TST
until…
DIAG ELEC
PMT = 1732 MV
NOX=X.X
<TST TST> EXIT
While the electrical test is
activated,
PMT
should equal:
2000 mV ± 1000 mV
SETUP X.X
ENTER DIAG PASS: 818
8 1 8 ENTR EXIT
04521C (DCN5731)