6.13.3. ANALOG OUTPUT STEP TEST
This test can be used to check the accuracy and proper operation of the analog outputs. The test forces all four
analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20% increments.
This test is useful to verify the operation of the data logging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit-Exit
returns to the
DIAG menu
SETUP X.X ENTER DIAG PASS: 818
8 1 8 ENTR EXIT
DIAG ANALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG AOUT ANALOG OUTPUT
0% EXIT
Performs
analog output
step test.
0% - 100%
DIAG AOUT ANALOG OUTPUT
[0%] EXIT
Pressing the key under “0%” while performing the test will
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again
will resume the test.