Teledyne API - Model 200EH/EM Operation Manual Operating Instructions
147
To Initialize the modem press:
Select which
COM Port is
tested
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X COMMUNICATIONS MENU
ID COM1 COM2 EXIT
EXIT returns
to the
previous
menu
SETUP X.X COM1 MODEM INIT:AT Y &D
&H
<SET SET> EDIT EXIT
SETUP X.X COM1 INITIALIZE MODEM
<SET SET> INIT EXIT
SETUP X.X COM1 MODE:0
SET> EDIT EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET SET> EDIT EXIT
EXIT returns to the
Communications Menu.
SETUP X.X INITIALIZING MODEM
<SET SET> INIT EXIT
6.15.2.7. COM Port Password Security
In order to provide security for remote access of the M200EH/EM, a LOGON feature can be enabled to require a
password before the instrument will accept commands. This is done by turning on the
SECURITY MODE (see
Section 6.11.8). Once the
SECURITY MODE is enabled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the
LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be
disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.
The following messages will be returned at logon:
o LOGON SUCCESSFUL - Correct password given
o LOGON FAILED - Password not given or incorrect
o LOGOFF SUCCESSFUL - Connection terminated successfully
To log on to the M200EH/EM analyzer with
SECURITY MODE feature enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable RS232_PASS issued as
follows:
V RS232_PASS=NNNNNN
Where N is any numeral between 0 and 9.
04521C (DCN5731)