Teledyne API - Model 200EH/EM Operation Manual Operating Instructions
131
6.13.6.4. AIN Calibration
This is the sub-menu calibrates the analyzer’s A-to-D conversion circuitry. This calibration should only be
necessary after major repair such as a replacement of CPU, motherboard or power supplies.
To perform a
AIN CALIBRATION, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
Continue pressing SET> until ….
DIAG AIO AIN CALIBRATED: NO
<SET CAL EXIT
DIAG AIO CALIBRATING A/D ZERO
DIAG AIO CALIBRATING A/D SPAN
DIAG AIO AIN CALIBRATED: YES
<SET CAL EXIT
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT