6.13.5.6. Setting the Reporting Range Scale for an Analog Output
Once the iDAS parameter has been set, the top end of the scale must be selected. For concentration values
this should be equal to the expected maximum value for the application. The analog channel will scale its output
accordingly.
EXAMPLE:
IDAS parameter being output: NXCNC1
Maximum value expected: 800 ppm
Output range; 10 V
Output: 0 ppm......... 0.000 V
100 ppm......1.250 V
200 ppm......2.500 V
400 ppm......5.000 V
750 ppm......9.375 V
NOTE
Regardless of how the reporting range for an analog output channel is set, the
instrument will continue to measure NO, NO
2
and NO
x
accurately for the entire physical
range of the instrument (See Section 6.13.3.2 for information on Physical range versus
reporting range)
Each output channel can be programmed for a separate gas with independent reporting range.
EXAMPLE:
A1 NXCNC1 (NOx Range 1) 0-1000 ppm NOX.
A1 NXCNC2 (NOx Range 2) 0-1250 ppm NOX.
A3 NOCNC1 (NOx Range 1) 0-500 ppm NO.
A4 N2CNC1 (NO2 Range 1) 0-750 ppm NO
2
.
NOTE:
While Range 1 for each gas type is often referred to as the LOW range and Range 2 as
the HIGH range, this is simply a naming convention. The upper limit for each range can
be set to any value.
EXAMPLE: A1 NXCNC1 (NOx Range 1) 0-1500 ppm NOX
A2 NXCNC2 (NOx Range 2) 0-1000 ppm NOX.