7. Calibrating the output channel. This can be done automatically or manually for each channel (see
Sections 6.13.5).
To access the analog I/O configuration sub menu, press:
AIO Configuration Submenu
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTREXIT
DIAG SIGNAL I/O
NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO A OUTS CALIBRATED: NO
<SET SET> CAL EXIT
DIAG AIO DATA_OUT_1: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO AIN CALIBRATED: NO
<SET SET> CAL EXIT
DIAG AIO DATA_OUT_2: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_4: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT