Appendix C
(Ref: 05149A)
Warranty/Repair
Questionnaire Model
200EH/EM
TELEDYNE
INSTRUMENTS
Advanced Pollution Instrumentation
A Teledyne Technologies Company
TELEDYNE INSTRUMENTS CUSTOMER SERVICE
EMAIL: api-customerservice@teledyne.com
PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816
C-1
CUSTOMER:_____________________________________ PHONE: ________________________________
CONTACT NAME: ________________________________ FAX NO. _______________________________
SITE ADDRESS:_____________________________________________________________________________
MODEL TYPE: ______________ SERIAL NO.:_________________ FIRMWARE REVISION: ___________
1. Are there any failure messages? ______________________________________________________________
__________________________________________________________________________________________
__________________________________________________________________________________________
________________________________________________________________ (Continue on back if necessary)
PLEASE COMPLETE THE FOLLOWING TABLE:
TEST FUNCTION RECORDED VALUE UNITS ACCEPTABLE VALUE
NOx STAB PPB/PPM
1 PPB WITH ZERO AIR
SAMPLE FLOW
CM
3
500 ± 50
OZONE FLOW
CM
3
80 ± 15
PMT SIGNAL WITH ZERO AIR MV
-20 to 150
PMT SIGNAL AT SPAN GAS CONC
MV
PPB
0-5000MV
0-5,000 PPM
1
, 200 PPM
2
NORM PMT SIGNAL AT SPAN
GAS CONC
MV
PPB
0-5000MV
0-5,000 PPM
1
, 200 PPM
2
AZERO MV
-20 to 150
HVPS V
400 to 900
RCELL TEMP ºC
50 ± 1
BOX TEMP ºC
AMBIENT ± 5ºC
PMT TEMP ºC
7 ± 2ºC
O
2
CELL TEMP
3
ºC
30ºC to 70ºC
IZS TEMP
3
ºC
50 ± 1ºC
MOLY TEMP ºC
315 ± 5ºC
RCEL IN-HG-A
<10
SAMP IN-HG-A
AMBIENT ± 1
NOx SLOPE
1.0 ± 0.3
NOx OFFSET mV
50 to 150
NO SLOPE
1.0 ± 0.3
NO OFFSET mV
50 to 150
O2 SLOPE
3
0.5 to 2.0
O2 OFFSET
3
%
-10 to + 10
PMT SIGNAL DURING ETEST MV
2000 ± 1000
PMT SIGNAL DURING OTEST MV
2000 ± 1000
REF_4096_MV
4
MV
4096mv ±2mv and Must be
Stable
REF_GND
4
MV
0± 0.5 and Must be Stable
1
M200EH
2
M200EM
3
If option is installed
4
Located in Signal I/O list under DIAG menu
04521C (DCN5731)