6.13. SETUP MORE DIAG: DIAGNOSTICS MENU
A series of diagnostic tools is grouped together under the SETUP-MORE-DIAG menu. These parameters are
dependent on firmware revision. These tools can be used in a variety of troubleshooting and diagnostic
procedures and are referred to in many places of the maintenance and trouble-shooting sections.
An overview of the entire
DIAG menu can be found in menu tree A-6 of Appendix A.1.
Table 6-16: M200EH/EM Diagnostic (DIAG) Functions
DIAGNOSTIC FUNCTION AND MEANING
FRONT PANEL
MODE
INDICATOR
SECTION
SIGNAL I/O: Allows observation of all digital and analog signals in the
instrument. Allows certain digital signals such as valves and heaters to be
toggled ON and OFF.
DIAG I/O
6.13.2
ANALOG I/O: When entered, the analyzer performs an analog output step test.
This can be used to calibrate a chart recorder or to test the analog output
accuracy.
DIAG AOUT
6.13.3
ANALOG I/O CONFIGURATION: This submenu allows the user to configure
the analyzer’s four analog output channels, including choosing what parameter
will be output on each channel. Instructions that appear here allow adjustment
and calibration the voltage signals associated with each output as well as
calibration of the analog to digital converter circuitry on the motherboard.
DIAG AIO
6.13.4,
through
6.13.6
DISPLAY SEQUENCE CONFIGURATION: Allows the user to program which
concentration values are displayed in the .
DIAG DISP
6.13.7.1
OPTIC TEST: When activated, the analyzer performs an optic test, which turns
on an LED located inside the sensor module near the PMT (Fig. 10-15). This
diagnostic tests the response of the PMT without having to supply span gas.
DIAG OPTIC
6.13.7.2
ELECTRICAL TEST: When activated, the analyzer performs an electric test,
which generates a current intended to simulate the PMT output to verify the
signal handling and conditioning of the PMT preamp board.
DIAG ELEC
6.13.7.3
OZONE GEN OVERRIDE: Allows the user to manually turn the O
3
generator on
or off. This setting is retained when exiting
DIAG.
DIAG OZONE
6.13.7.4
FLOW CALIBRATION: This function is used to calibrate the gas flow output
signals of sample gas and ozone supply. These settings are retained when
exiting
DIAG.
DIAG FCAL
6.13.7.5
04521C (DCN5731)