Operating Instructions Teledyne API - Model 200EH/EM Operation Manual
132
6.13.7. OTHER DIAG MENU FUNCTIONS
6.13.7.1. Display Sequence Configuration
The model M200EH/EM analyzer allows the user to choose which gas concentration measurement and
reporting range is to be displayed in the concentration field on the instrument’s front panel display as well as
what order and how long each will appear before analyzer cycle to the next item on the display list.
NOTE
This M200EH/EM is constantly monitoring all of the gas measurements it is configured
to make regardless of which range is being displayed. This feature merely changes how
that display sequence occurs, not how the instrument makes measurements.
The software permits the user to choose from the following list of display values:
Table 6-26: M200EH/EM Available Concentration Display Values
DISPLAY
VALUE
DESCRIPTION
ASSOCIATED iDAS
PARAMETER
NOX
NO
x
value computed with the slope and offset values for the currently selected
NO
x
range.
1
--
NXL
NO
x
value computed with the slope and offset values for NO
x
reporting range
1 (Low)
NXCNC1
NXH
NO
x
value computed with the slope and offset values for NO
x
reporting range
2 (High)
NXCNC2
NO
NO value of computed with the slope and offset values for the currently
selected NO range
1
--
NOL
NO value computed with the slope and offset values for NO reporting range 1
(Low)
NOCNC1
NOH
NO value computed with the slope and offset values for NO reporting range 2
(High)
NOCNC2
N2
NO
2
value of computed with the slope and offset values for the currently
selected NO
2
range
1
--
N2L
NO
2
value computed for with the slope and offset values for NO
x
reporting
range 1 (Low) & N0 reporting range 1 (Low)
N2CNC1
N2H
NO
2
value computed for with the slope and offset values for NO
x
reporting
range 2 (High) & N0 reporting range 2 (High)
N2CNC2
O2
O
2
concentration value.
O2CONC
2
1
With the following exceptions this will be reporting range 1 (Low) for the appropriate gas type:
If the analyzer is in calibration mode, this will be the concentration value computed with the slope and offset for which ever range
is being calibrated.
If the instrument is in either E-Test or O-Test mode, this will be the value computed with the slope and offset values used by
those tests.
2
Only appears if O2 sensor option is installed.
04521C (DCN5731)