Teledyne 200EH/EM TV Converter Box User Manual


 
Operating Instructions Teledyne API - Model 200EH/EM Operation Manual
154
To assign or reset the status flag bit assignments, press:
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X. HESSEN STATUS FLAGS
<SET SET> EDIT EXIT
SETUP X.X COMMUNICATIONS MENU
ID HESN COM1 COM2 EXIT
SETUP X. PMT DET WARNING: 0002
PREV NEXT EDIT PRNT EXIT
Repeat pressing SET> until …
SETUP X. SYSTEM RESET: 0000
PREV NEXT EDIT PRNT EXIT
The <CH and
CH> keys move
the [ ] cursor left
and right along
the bit string.
Repeat pressing NEXT or PREV until the desired
message flag is displayed. See Table 6-29.
For example …
ENTR key accepts the
new settings
EXIT key ignores the new
settings
SETUP X. SYSTEM RESET: [0]000
<CH CH> [0] ENTR EXIT
Press the [?] key repeatedly to cycle through the available character set: 0-9
Note: Values of A-F can also be set but are meaningless.
6.15.4.7. Instrument ID Code
Each instrument on a Hessen Protocol network must have a unique ID code. The M200EH/EM is programmed
with a default ID code of
200. To change this code see Section 6.11.1