6.13.5.3. Turning the Analog Output Over-Range Feature ON/OFF
In its default configuration a ± 5% over-range is available on each of the M200EH/EM’s analog output channels.
This over-range can be disabled if your recording device is sensitive to excess voltage or current.
NOTE:
Instruments with current range options installed on one or more of the outputs often
are delivered from the factory with the over-range feature turned OFF on those
channels.
To Turn the over-range feature on or off, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_2 OVERRANGE: ON
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_2 RANGE: 5V
SET> EDIT EXIT
DIAG AIO DATA_OUT_2 OVERRANGE: ON
ON ENTR EXIT
DIAG AIO DATA_OUT_2 OVERRANGE: OFF
OFF ENTR EXIT
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
Toggle this key
to turn the
Over-Range
feature
ON/OFF
Continue pressing SET> until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT