Teledyne 200EH/EM TV Converter Box User Manual


 
Theory of Operation Teledyne API - Model 200EH/EM Operation Manual
216
10.4.7.3. Thermocouple Inputs and Configuration Jumper (JP5)
Although the relay PCA supports two thermocouple inputs, the current M200EH/EM series analyzers only utilize
one. By default, this single thermocouple input is plugged into the TC1 input (J15). TC2 (J16) is currently not
used. See Figure 11-4 for location of J15 and J16
Table 10-4: Thermocouple Configuration Jumper (JP5) Pin-Outs
TC INPUT JUMPER PAIR DESCRIPTION FUNCTION
1 – 11
Gain Selector
Selects preamp gain factor for J or K TC
- IN = J TC gain factor
- OUT = K TC gain factor
2 – 12
Output Scale Selector
Selects preamp gain factor for J or K TC
- IN = 5 mV / °C
- OUT = 10 mV / °C
3 – 13
Type J Compensation
When present, sets Cold Junction
Compensation for J type Thermocouple
4 – 14
Type K Compensation
When present, sets Cold Junction
Compensation for K type Thermocouple
TC1
5 – 15
Termination Selector
Selects between Isolated and grounded TC
- IN =
Isolate TC
- OUT =
Grounded TC
6 – 16
Gain Selector
Same as Pins 1 – 11 above.
7 – 17
Output Scale Selector
Same as Pins 2 – 12 above.
8 – 18
Type J Compensation
Same as Pins 3 – 13 above.
9 – 19
Type K Compensation
Same as Pins 4 – 14 above.
TC2
10 – 20
Termination Selector
Same as Pins 5 – 15 above.
TC2
TC1
Input Gain Selector 1 – 11
Type J Compensation 4 – 14
Output Scale Selector 2 – 12
Type J Compensation 3 – 13
Termination Selector 5 – 15
Input Gain Selector 6 – 16
Type J Compensation 9 – 19
Output Scale Selector 7 – 17
Type J Compensation 8 – 18
Termination Selector 10 – 20
Figure 10-10-23: Thermocouple Configuration Jumper (JP5) Pin-Outs
04521C (DCN5731)