Teledyne API - Model 200EH/EM Operation Manual Operating Instructions
117
6.13.5.4. Adding a Recorder Offset to an Analog Output
Some analog signal recorders require that the zero signal is significantly different from the baseline of the
recorder in order to record slightly negative readings from noise around the zero point. This can be achieved in
the M200EH/EM by defining a zero offset, a small voltage (e.g., 10% of span).
To add a zero offset to a specific analog output channel, press:
EXAMPLE
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_2 REC OFS: 0 mV
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_2 OUTPUT: 5V
SET> EDIT EXIT
DIAG AIO DATA_OUT_2 REC OFS: 0 mV
+ 000 0ENTREXIT
DIAG AIO DATA_OUT_2 REC OFS: -10 mV
<SET SET> EDIT EXIT
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
Toggle these
keys to set ther
value of the
desired offset.
Continue pressing SET> until ...
DIAG AIO DATA_OUT_2 REC OFS: -10 mV
– 001 0 ENTR EXIT
Continue pressing SET> until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT
04521C (DCN5731)