6.13.7.4. Ozone Generator Override
This feature allows the user to manually turn the ozone generator off and on. This can be done before
disconnecting the generator, to prevent ozone from leaking out, or after a system restart if the user does not
want to wait for 30 minutes during warm-up time. Note that this is one of the two settings in the
DIAG menu that
is retained after you exit the menu. To access this feature press the following keys. Also note that the ozone
generator does not turn on if the ozone flow conditions are out of specification (e.g., if there is no flow through
the system or the pump is broken).
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X ENTER DIAG PASS: 818
8 1 8 ENTR EXIT
DIAG SIGNAL I / O
PREV NEXT JUMP ENTR EXIT
DIAG OZONE
OZONE GEN OVERRIDE
OFF ENTR EXIT
Press NEXT until…
DIAG OZONE GEN OVERRIDE
PREV NEXT ENTR EXIT
Toggle this key to turn the O
3
generator ON/OFF.