Troubleshooting & Repair Teledyne API - Model 200EH/EM Operation Manual
254
The Relay PCA 04523, includes a set of jumpers that connect AC power to heaters included in several
optional items, such as the zero/span valve options and the O
2
sensor option available on the
M200EH/EM analyzers. In earlier versions of the relay PCA this was also handled by in-line
connections.
11.5.4.1. AC configuration – Internal Pump (JP7)
AC power configuration for internal pumps is set using Jumper set JP7 (see Figure 11-4 for the location of JP7).
Table 11-3: AC Power Configuration for Internal Pumps (JP7)
LINE
POWER
LINE
FREQUENCY
JUMPER
COLOR
FUNCTION
JUMPER
BETWEEN
PINS
Connects pump pin 3 to 110 / 115 VAC power line 2 to 7
Connects pump pin 3 to 110 / 115 VAC power line 3 to 8
60 HZ WHITE
Connects pump pins 2 & 4 to Neutral 4 to 9
Connects pump pin 3 to 110 / 115 VAC power line 2 to 7
Connects pump pin 3 to 110 / 115 VAC power line 3 to 8
110VAC
115 VAC
50 HZ
1
BLACK
Connects pump pins 2 & 4 to Neutral 4 to 9
Connects pump pins 3 and 4 together 1 to 6
60 HZ BROWN
Connects pump pin 1 to 220 / 240VAC power line 3 to 8
Connects pump pins 3 and 4 together 1 to 6
220VAC
240 VAC
50 HZ
1
BLUE
Connects pump pin 1 to 220 / 240VAC power line 3 to 8
1
A jumper between pins 5 and 10 may be present on the jumper plug assembly, but is only functional on the M300E and
has no function on the M200EH/EM analyzers.
110 VAC /115 VAC
220 VAC /240 VAC
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
Present on 50 Hz version of jumper set,
and functional for M300E but not
Models M100E, M200E & M400E
Figure 11-16: Pump AC Power Jumpers (JP7)
04521C (DCN5731)