Teledyne 200EH/EM TV Converter Box User Manual


 
Model 200EH/EM (Ref: 05147F) APPENDIX A-3: Warnings and Test Functions, Revision F.0
A-17
Table A-3: M200EH/EM Test Functions, Revision F.0
TEST Function Message Text Description
RNG_DATA_OUT_1 “A1:<data point>=<range> <units>” D/A #1 range.
RNG_DATA_OUT_2 “A2:<data point>=<range> <units>” D/A #2 range.
RNG_DATA_OUT_3 “A3:<data point>=<range> <units>” D/A #3 range.
RNG_DATA_OUT_4 “A4:<data point>=<range> <units>” D/A #4 range.
NONOXCONC
1
NO=396.5 NOX=396.5
3
Simultaneously displays NO and NOX
concentrations.
RANGE RANGE=500.0 PPB
3
D/A range in single or auto-range modes.
RANGE1 RANGE1=500.0 PPB
3
D/A #1 range in independent range mode.
RANGE2 RANGE2=500.0 PPB
3
D/A #2 range in independent range mode.
RANGE3 RANGE3=500.0 PPB
3
D/A #3 range in independent range mode.
O2RANGE
2
O2 RANGE=200.00 % D/A #4 range for O
2
concentration.
STABILITY NOX STB=0.0 PPB
3
Concentration stability (standard deviation
based on setting of STABIL_FREQ and
STABIL_SAMPLES). Select gas with
STABIL_GAS variable.
RESPONSE
2
RSP=8.81(1.30) SEC
Instrument response. Length of each signal
processing loop. Time in parenthesis is
standard deviation.
SAMPFLOW SAMP FLW=460 CC/M Sample flow rate.
OZONEFLOW OZONE FL=87 CC/M Ozone flow rate.
PMT PMT=800.0 MV Raw PMT reading.
NORMPMT NORM PMT=793.0 MV
PMT reading normalized for temperature,
pressure, auto-zero offset, but not range.
AUTOZERO AZERO=1.3 MV Auto-zero offset.
HVPS HVPS=650 V High voltage power supply output.
RCELLTEMP RCELL TEMP=50.8 C Reaction cell temperature.
BOXTEMP BOX TEMP=28.2 C Internal chassis temperature.
PMTTEMP PMT TEMP=7.0 C PMT temperature.
MANIFOLDTEMP MF TEMP=50.8 C Bypass or dilution manifold temperature.
O2CELLTEMP
2
O2 CELL TEMP=50.8 C O
2
sensor cell temperature.
IZSTEMP IZS TEMP=50.8 C IZS temperature.
CONVTEMP MOLY TEMP=315.0 C
Converter temperature. Converter type is
MOLY, CONV, or O3KL.
RCELLPRESS RCEL=7.0 IN-HG-A Reaction cell pressure.
SAMPPRESS SAMP=29.9 IN-HG-A Sample pressure.
NOXSLOPE NOX SLOPE=1.000
NOX slope for current range, computed during
zero/span calibration.
NOXOFFSET NOX OFFS=0.0 MV
NOX offset for current range, computed during
zero/span calibration.
NOSLOPE NO SLOPE=1.000
NO slope for current range, computed during
zero/span calibration.
NOOFFSET NO OFFS=0.0 MV
NO offset for current range, computed during
zero/span calibration.
NO
2
NO2=0.0 PPB
3
NO
2
concentration for current range.
NOX NOX=396.5 PPB
3
NOX concentration for current range.
04521C (DCN5731)