Operating Instructions Teledyne API - Model 200EH/EM Operation Manual
128
To manually adjust the signal levels of an analog output channel, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
Continue pressing SET> until you reach the
output to be configured
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
Continue pressing SET> until ...
DIAG AIO DATA_OUT_2 CALIBRATED:NO
<SET SET> CAL EXIT
DIAG AIO DATA_OUT_2 RANGE: 5V
SET> EDIT EXIT
DIAG AIO DATA_OUT_2 VOLT-Z: 0 mV
U100 UP10 UP DOWN DN10 D100 ENTREXIT
DIAG AIO DATA_OUT_2 VOLT-S: 4500 mV
U100 UP10 UP DOWN DN10 D100 ENTREXIT
DIAG AIO DATA_OUT_2 CALIBRATED: YES
<SET SET> CAL EXIT
These menu’s
only appear if
AUTO-CAL is
turned OFF
These keys increase / decrease
the analog output signal level
(not the value on the display)
by 100, 10 or 1 counts.
Continue adjustments until the
voltage measured at the output
of the analyzer and/or the input
of the recording device matches
the value in the upper right hand
corner of the display (within the
tolerances
listed in Table 6-24.
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT
6.13.6.3. Manual Calibration of Analog Outputs configured for Current Loop Ranges
The current loop output option (see Section 5.4) uses a small converter assembly to change the DC voltage
output by the standard voltage output to a current signal ranging between 0-20 mA. Since the exact current
increment per voltage count varies from converter to converter and from instrument to instrument, analog
outputs with this option installed cannot be calibrated automatically and must be adjusted manually.
Adjusting the signal zero and full scale values of the current loop output is done in a similar manner as manually
adjusting analog outputs configured for voltage output except that:
In this case calibration is performed with a current meter connected in series with the output circuitry
(See Figure 6-6-15).
Adjustments to the output are made usi
ng the front panel keys, also in 100, 10 or 1 count increments,
but the change in the voltage driving the converter assembly is displayed on the front panel.
As before, adjustment of the output is performed until the current reading of the meter reaches the
desired point (e.g. 2 mA, 4 mA, 20 mA, etc.)
04521C (DCN5731)