Teledyne 200EH/EM TV Converter Box User Manual


 
Operating Instructions Teledyne API - Model 200EH/EM Operation Manual
110
6.13.4. ANALOG OUTPUTS AND REPORTING RANGES
6.13.4.1. Analog Output Signals Available on the M200EH/EM
The analyzer has four analog output signals, accessible through a connector on the rear panel.
ANALOG OUT
A1
A
2
A
3
A
4
+ - + - + - + -
0-20 mA current loop
output available for these
channels only
Figure 6-6-13: Analog Output Connector Key
The signal levels of each output can be independently configured as follows. An over-range feature is available
that allows each range to be usable from -5% to + 5% of its nominal scale:
Table 6-17: Analog Output Voltage Ranges with Over-Range Active
RANGE MINIMUM OUTPUT MAXIMUM OUTPUT
0-0.1 V -5 mV +105 mV
0-1 V -0.05 V +1.05 V
0-5 V -0.25 V +5.25 V
0-10 V -0.5 V +10.5 V
The default offset for all ranges is 0 VDC.
Pin assignments for the ANALOG output connector at the rear panel of the instrument:
Table 6-18: Analog Output Pin Assignments
PIN
ANALOG
OUTPUT
VOLTAGE
SIGNAL
CURRENT
SIGNAL
1 V Out I Out +
2
A1
Ground I Out -
3 V Out I Out +
4
A2
Ground I Out -
5 V Out I Out +
6
A3
Ground I Out -
7 V Out I Out +
8
A4
Ground I Out -
04521C (DCN5731)