6.13.2. SIGNAL I/O
The signal I/O diagnostic mode allows to review and change the digital and analog input/output functions of the
analyzer. See Appendix A-4 for a complete list of the parameters available for review under this menu.
NOTE
Any changes of signal I/O settings will remain in effect only until the signal I/O menu
is exited. Exceptions are the ozone generator override and the flow sensor calibration,
which remain as entered when exiting.
To enter the signal I/O test mode, press:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
EXAMPLE
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG EXIT
DIAG
SIGNAL I / O
PREV NEXT JUMP ENTR EXIT
DIAG I / O
Test Signals Displayed Here
PREV NEXT JUMP PRNT EXIT
Use the JUMP key to
go directly to a
specific signal
See Appendix A-4 for
a complete list of
available SIGNALS
DIAG I / O JUMP TO: 5
0 5 ENTR
EXIT
Enter 05 to Jump
to Signal 5:
(
CAL_LED
)
DIAG I / O CAL_LED = ON
PREV NEXT JUMP ON PRNT EXIT
Exit to return
to the
DIAG
menu
SETUP X.X ENTER DIAG PASS: 818
8 1 8 ENTR EXIT
Use the
NEXT
&
PREV
keys to move between
signal types.
Pressing the PRNT key will send a formatted printout to the serial port and can be
ca
p
tured with a com
p
uter or other out
p
ut device.
EXIT
returns
to the main
SAMPLE
display