3.1.3.2. Pneumatic Connections to M200EH/EM Basic Configuration:
Figures 3-7 and 3-8 show the most common configurations for gas supply and exhaust lines to the Model
200EH/EM Analyzer. Please refer to Figure 3-2 for the locations of pneumatic connections on the rear panel and
Table 3-5 for nomenclature.
NOTE
Sample and calibration gases should only come into contact with PTFE (Teflon) or glass
or materials. They should not come in contact with FEP or stainless steel materials.
Source of
SAMPLE GAS
Removed during
calibration
MODEL
200EH/EM
SAMPLE
EXHAUST
PUMP
MODEL 700
Gas Dilution
Calibrator
VENT
MODEL 701
Zero Gas
Generator
NO
x
Gas
(High Concentration)
VENT here if input
is pressurized
Figure 3-9: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator
Source of
SAMPLE GAS
Removed during
calibration
MODEL
200EH/EM
SAMPLE
EXHAUST
PUMP
MODEL 701
Zero Gas
Generator
VENT
3-way Valve
Manual
Control Valve
VENT here if input
is pressurized
NO
X
Gas
(High Concentration)
Figure 3-10: Pneumatic Connections–Basic Configuration–Using Bottled Span Gas