Teledyne 200EH/EM TV Converter Box User Manual


 
Teledyne API - Model 200EH/EM Operation Manual Operating Instructions
71
6.7.2.4. Editing iDAS Parameters
Data channels can be edited individually from the front panel without affecting other data channels. However,
when editing a data channel, such as during adding, deleting or editing parameters, all data for that particular
channel will be lost, because the iDAS can store only data of one format (number of parameter columns etc.) for
any given channel. In addition, an iDAS configuration can only be uploaded remotely as an entire set of
channels. Hence, remote update of the iDAS will always delete all current channels and stored data.
To modify, add or delete a parameter, follow the instruction shown in section 6.7.2.2 then press:
Edit Data Parameter Menu
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 8, 800
PREV NEXT INS DEL
EDIT
PRNT
EXIT
From the
DATA ACQUISITION
menu
(see Section 6.7.2.2)
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X
PARAMETERS:
8
<SET SET>
EDIT
PRINT EXIT
SETUP X.X
0) PARAM=DETREP, MODE=INST
PREV NEXT
INS DEL EDIT EXIT
Press
SET>
key until…
SETUP X.X
EDIT PARAMS (DELETE DATA)
Y
ES NO
NO returns to
the previous
menu and
retains all data.
Moves the
display between
available
Parameters
Inserts a new Parameter
before the currently
displayed Parameter
Deletes the Parameter
currently displayed.
Use to configure
the functions for
this Parameter.
Exits to the main
Data Acquisition
menu
YES will delete
all data in that
entire channel.
04521C (DCN5731)