6.12. SETUP MORE VARS: INTERNAL VARIABLES (VARS)
The M200EH/EM has several-user adjustable software variables, which define certain operational parameters.
Usually, these variables are automatically set by the instrument’s firmware, but can be manually re-defined using
the VARS menu. Table 6-15 lists all variables that are available wi
thin the 818 password protected level. See
Appendix A2
for a detailed listing of all of the M200EH/EM variables that are accessible through the remote
interface.
Table 6-15: Variable Names (VARS)
NO. VARIABLE DESCRIPTION ALLOWED VALUES
0
DAS_HOLD_OFF
Duration of no data storage in the iDAS. This is the time when
the analyzer returns from one of its calibration modes to the
SAMPLE mode. The DAS_HOLD_OFF can be disabled in each
iDAS channel.
Can be between 0.5
and 20 minutes
Default=15 min.
1
MEASURE_MODE
Selects the gas measurement mode in which the instrument is to
operate. NO
x
only, NO only or dual gas measurement of NO
x
and NO simultaneously. Dual gas mode requires that a special
switching optional be installed.
NO; NO
x
;
NO
x
–NO
2
STABIL_GAS
Selects which gas measurement is displayed when the STABIL
test function is selected.
NO; NO
x
;
NO
2
; O
2
1
3
TPC_ENABLE
Enables or disables the temperature and pressure
compensation (TPC) feature (Section 10.7.3).
ON/OFF
Default=ON
4
DYN_ZERO
Dynamic zero automatically adjusts offset and slope of the NO
and NO
X
response when performing a zero point calibration
during an AutoCal (Chapter 7.7).
ON/OFF
Default=OFF
5
DYN_SPAN
Dynamic span automatically adjusts the offsets and slopes of
the NO and NOx response when performing a zero point
calibration during an AutoCal (Chapter 7.7).
Note that the DYN_ZERO and DYN_SPAN features are not
allowed for applications requiring EPA equivalency.
ON/OFF
Default=OFF
6
CONC_PRECISION
Allows to set the number of decimal points of the concentration
and stability parameters displayed on the front panel.
AUTO, 1, 2, 3, 4
Default=AUTO
7
CLOCK_ADJ
Adjusts the speed of the analyzer’s clock. Choose the + sign if
the clock is too slow, choose the - sign if the clock is too fast.
-60 to +60 s/day
Default=0
1
Only available in analyzer’s with o
2
sensor options installed.
NOTE:
There is a 2 second latency period between when a VARS value is changed and the new
value is stored into the analyzer’s memory.
DO NOT turn the analyzer off during this period or the new setting will be lost.